EA015 MULTI FUNCTION WORKSTATION
OPERATION MANUAL
Transmille Ltd. Page 16
Note : Only the outputs which are available for current value entered are
enabled – if the button is not available (greyed out) this means it is not
possible to output this value using that specific output (eg. It is not possible to
output 2A using the 50 turn coil – minimum value from 50 turn coil is 100A :
2A x 50 turns).
STEP 5 : Set Output
Once the output value is set, press the enter button to set the
output.
Use the STANDBY and OUTPUT ON buttons to control the
output.
The Autoset Output On option allows the output to be
automatically switched on when a value is entered. If this option is unchecked, the
output will not be switch on automatically, but left set in standby.
When high voltage (>40V) is output, the
Autoset Output ON is not available.
Changing Output Values
If the output value is changed, the VFP program will check if this is possible using
the current range selected. If it is not possible to output this value, the current output
buttons will flash to indicate that a different output should be selected.
Note : Only the outputs which are available for current value entered are enabled,
e.g.
Example : 1000A Output Set
50 TURN COIL (to 1500A) enabled,
all other modes disabled as not
possible with 1000A Selected.
10 TURN COIL – Max. Output = 300A
2 TURN COIL – Max. Output = 60A
TERMINALS
Max. Output
=
30A
Page view 16